Series 3000 Trace Oxygen Analyzer
Series 3000 Trace Oxygen Analyzer
Our ALPHA OMEGA INSTRUMENTS™ Series 3000™ Trace Oxygen Analyzer is a rugged, microprocessor-controlled oxygen analyzer designed for continuous or intermittent measurement of trace oxygen concentrations.
Pricing for US, LATAM, and Canada Only.
Outside of US, LATAM, or Canada, please request a quote. sales.hou@process-insights.com
The Series 3000 Trace Oxygen Analyzer is housed in a general-purpose enclosure and is powered from either a universal 100-240 VAC, 50-60Hz, or 18-32 VDC. Trace oxygen values are displayed on a 0.4 inch (10.2 mm) high, 4-1/2-digit liquid crystal display (LCD). The Series 3000 is equipped with three oxygen alarm relays and one status alarm relay. All four relays are Form C (SPDT) types rated at 10 amps at 115/230 VAC and 30 VDC. The relays are user configurable for fail-safe operation. In addition to the four alarm contacts, the analyzer has a built-in audible alarm and three red LEDs for visual indication of an oxygen alarm condition.
The audible alarm may be manually canceled at any time. The Series 3000 provides users with two standard analog outputs, 4-20 mADC and 0-2 VDC. For enhanced communications, the analyzer can be equipped with optional RS-232C or RS-485 serial communications. Also available with an optional CO2 tolerant sensor.
NEW CO2 RESISTANT SENSOR
We offer an optional CO2 tolerant trace oxygen sensor with proprietary electrolyte. The CO2 tolerant sensor is capable of providing accurate oxygen readings in gases containing up to 100% CO2 without shortening the life of the sensor.
FEATURES
- Long-life Trace Oxygen Sensor: Provides up to three times the functional life of most “fuel cell” type sensors. Reliability is dramatically increased while maintenance costs are reduced.
- Sealed Sensor: Eliminates handling potassium hydroxide (caustic), a hazardous material or having to frequently “recharge” the electrolyte.
- CO2 Resistant Sensor: Optional CO2 resistant sensor is designed to operate with sample gases containing up to 100% CO2. Conventional “fuel cell” type sensors using KOH electrolyte can be poisoned from CO2.
- Affordable: Economically priced with savings of up to 50% over other manufacturers trace oxygen analyzers.
- Wide Measurement Range: Measure from under 1 PPM to as high as 10,000 PPM oxygen.
- Factory Calibration: A factory Certificate of Performance Verification and Calibration is included at no additional charge. The product performance is verified in accordance with manufacturing specifications using a National Institute of Standards and Technology (NIST) traceable transfer calibrated standard gas mix.
- Visual and audible alarms included along with two analog recorder outputs.
PERFORMANCE
- Ranges (parts per million): Single-range 0-10,000; 0- 5,000; 0-1,000; 0-100; 0- 50; Three-range 0-100/1,000/10,000; 0- 50/500/5,000; 0-10/100/1,000; 0- 5/50/500; 0-1/10/100
- Accuracy¹: ±1% of full scale (± 5% of full scale on ranges ≤ 0-5ppm)
- ¹Stated at constant temperature and constant pressure.
- Linearity: ±1%
- Response Time: 90% of full scale response in less than 10 seconds (typical). The response time for ranges of 0-50 PPM or less depend to a great extent on the design of the sample delivery system including the materials used.
- Sensor Type: Long-life Ambient Temperature Electrochemical Sensor(Optional CO2 Resistant Sensor Available).
- Temperature Compensation: Standard
- Operating Temperature: (both sensor and electronics): 40° to 104° F (5° to 40° C)
- Warranty: 2 years electronics, 1 year sensor
CONSTRUCTION
- Electronics Control Unit: Polycarbonate enclosure with a hinged clear front cover.
- Control Unit Dimensions: 9.45 in. (240.5 mm) height 6.50 in. (165.1 mm) width 6.20 inches (157.5 mm)
- Note: All dimensions are without optional equipment
- Gas Connections: 1/4″ stainless steel compression fittings.
- Sensor Mounting: Local
- Weight: 11 lbs (4.98 kg) (Standard Bench-top Configuration)
- Other mechanical configurations available upon request
ELECTRICAL
- Display: 0.4″ (10.2 mm) high, 4-1/2 digit LCD
- Resolution of Display: 0.1 ppm for ranges <100 ppm 1.0 ppm for ranges >100 ppm
- Input Power: Universal 100-240 VAC, 50-60Hz or 18-32 VDC.
- Outputs: 4-20 mADC and 0-2 VDC standard as is a TTL output for range identification. Optional RS-232C or RS-485 serial communication available.
- Oxygen Alarm Relays: Three (3) SPDT Form C contacts rated 10 A @ 30 VDC/115/230 VAC. User selectable for fail-safe or non fail-safe.
- Instrument Status Alarm: Rated identically to the O2 relays
- Audible Alarm: Internal buzzer with audible cancel
SAMPLE GAS CHARACTERISTICS
- Flow Rate: 1.0 to 2.0 SCFH (0.5 to 1.0 liter/min)
- Sample Gas Temperature: 40° to 104° F (5° to 40° C)
- Sample Gas Pressure: 0.1 to 1.5 psig (0.007 kg/cm² to 0.1 kg/cm²)
- Entrained Solids: <3 mg/ft³: no in-line filter required >3 mg/ft³: in-line filter is required
- Hydrocarbon Mist: <0.7 mg/ft³: no in-line filter required >0.7 mg/ft³: in-line filter is required
3000 with AC power 0-50/500/5000 PPM flowmeter regulator |
Series 3000 Trace Oxygen Analyzer with three alarm relays, instrument status alarm relay, digital display, audible alarm, visual alarm indicators, and two manual sensor isolation valves. "F" Auto Multi-range 0-50/500/5,000 PPM. Powered by universal AC (90-264VAC). Features a benchtop portable enclosure with carry handle. Includes standard connectors on the sample outlet; no filter; a pressure regulator (max 100 psi) with a stainless-steel needle valve with 1/4" compression fitting on the sample inlet; a flowmeter without a control valve on the sample outlet. Also includes a standard 4-20mADC and 0-2vdc analog outputs. Warranty: 2-year electronics and 1-year sensor. Pressure Regulator Application: For extractive gas sampling venting to atmosphere with sample gas pressure between 1.5 to 100 psig, flowrate 0.5 to 1.5 LPM. |
3000 with AC power 0-50/500/5000 PPM flowmeter pump |
Series 3000 Trace Oxygen Analyzer with three alarm relays, instrument status alarm relay, digital display, audible alarm, visual alarm indicators, and two manual sensor isolation valves. "F" Auto Multi-range 0-50/500/5,000 PPM. Powered by universal AC (90-264VAC). Features a benchtop portable enclosure with carry handle. Includes standard connectors on the sample outlet; no filter; a flowmeter with a control valve on the sample outlet; a 24vdc sample pump. Also includes a standard 4-20mADC and 0-2vdc analog outputs. Warranty: 2-year electronics and 1-year sensor. Sample Pump Application: For extractive gas sampling venting to atmosphere with sample gas pressure at atmosphere <0.1 psig, flowrate 0.5 to 1.5 LPM. Vacuums to 12.5 psia are permissible. For In-situ glovebox and chambers (instrument or sensor in chamber) with environment between atmosphere and 1.5 psig. Vacuums to 12.5 psia are permissible. |
3000 with AC power 0-50/500/5000 PPM no options |
Series 3000 Trace Oxygen Analyzer with three alarm relays, instrument status alarm relay, digital display, audible alarm, visual alarm indicators, and two manual sensor isolation valves. "F" Auto Multi-range 0-50/500/5,000 PPM. Powered by universal AC (90-264VAC). Features a benchtop portable enclosure with carry handle. Includes standard connectors on the sample outlet, no filter. Also includes a standard 4-20mADC and 0-2vdc analog outputs. Warranty: 2-year electronics and 1-year sensor. No options included. No Options Application: For extractive gas sampling venting to atmosphere with sample gas pressure between 0.1 to 1.5 psig, flowrate 0.5 to 1.5 LPM or customer will supply their own gas sample conditioning options. |
3000 with AC power 0-10/100/1000 PPM flowmeter regulator |
Series 3000 Trace Oxygen Analyzer with three alarm relays, instrument status alarm relay, digital display, audible alarm, visual alarm indicators, and two manual sensor isolation valves. "G" Auto Multi-range 0-10/100/1,000 PPM. Powered by universal AC (90-264VAC). Features a benchtop portable enclosure with carry handle. Includes standard connectors on the sample outlet; no filter; a pressure regulator (max 100 psi) with a stainless-steel needle valve with 1/4" compression fitting on the sample inlet; a flowmeter without a control valve on the sample outlet. Also includes a standard 4-20mADC and 0-2vdc analog outputs. Warranty: 2-year electronics and 1-year sensor. Pressure Regulator Application: For extractive gas sampling venting to atmosphere with sample gas pressure between 1.5 to 100 psig, flowrate 0.5 to 1.5 LPM. |
3000 with AC power 0-10/100/1000 PPM flowmeter pump |
Series 3000 Trace Oxygen Analyzer with three alarm relays, instrument status alarm relay, digital display, audible alarm, visual alarm indicators, and two manual sensor isolation valves. "G" Auto Multi-range 0-10/100/1,000 PPM. Powered by universal AC (90-264VAC). Features a benchtop portable enclosure with carry handle. Includes standard connectors on the sample outlet; no filter; a flowmeter with a control valve on the sample outlet; a 24vdc sample pump. Also includes a standard 4-20mADC and 0-2vdc analog outputs. Warranty: 2-year electronics and 1-year sensor Sample Pump Application: For extractive gas sampling venting to atmosphere with sample gas pressure at atmosphere <0.1 psig, flowrate 0.5 to 1.5 LPM. Vacuums to 12.5 psia are permissible. For In-situ glovebox and chambers (instrument or sensor in chamber) with environment between atmosphere and 1.5 psig. Vacuums to 12.5 psia are permissible. |
3000 with AC power 0-10/100/1000 PPM no options |
Series 3000 Trace Oxygen Analyzer with three alarm relays, instrument status alarm relay, digital display, audible alarm, visual alarm indicators, and two manual sensor isolation valves. "G" Auto Multi-range 0-10/100/1,000 PPM. Powered by universal AC (90-264VAC). Features a benchtop portable enclosure with carry handle. Includes standard connectors on the sample outlet, no filter. Also includes a standard 4-20mADC and 0-2vdc analog outputs. Warranty: 2-year electronics and 1-year sensor. No options included. No Options Application: For extractive gas sampling venting to atmosphere with sample gas pressure between 0.1 to 1.5 psig, flowrate 0.5 to 1.5 LPM or customer will supply their own gas sample conditioning options. |
3000 with AC power 0-100/1000/10000 PPM flowmeter regulator |
Series 3000 Trace Oxygen Analyzer with three alarm relays, instrument status alarm relay, digital display, audible alarm, visual alarm indicators, and two manual sensor isolation valves. "Y" Auto Multi-range 0-100/1,000/10,000 PPM. Powered by universal AC (90-264VAC). Features a benchtop portable enclosure with carry handle. Includes standard connectors on the sample outlet; no filter; a pressure regulator (max 100 psi) with a stainless-steel needle valve with 1/4" compression fitting on the sample inlet; a flowmeter without a control valve on the sample outlet. Also includes a standard 4-20mADC and 0-2vdc analog outputs. Warranty: 2-year electronics and 1-year sensor. Pressure Regulator Application: For extractive gas sampling venting to atmosphere with sample gas pressure between 1.5 to 100 psig, flowrate 0.5 to 1.5 LPM. |
3000 with AC power 0-100/1000/10000 PPM flowmeter pump |
Series 3000 Trace Oxygen Analyzer with three alarm relays, instrument status alarm relay, digital display, audible alarm, visual alarm indicators, and two manual sensor isolation valves. "Y" Auto Multi-range 0-100/1,000/10,000 PPM. Powered by universal AC (90-264VAC). Features a benchtop portable enclosure with carry handle. Includes standard connectors on the sample outlet; no filter; a flowmeter with a control valve on the sample outlet; a 24vdc sample pump. Also includes a standard 4-20mADC and 0-2vdc analog outputs. Warranty: 2-year electronics and 1-year sensor. Sample Pump Application: For extractive gas sampling venting to atmosphere with sample gas pressure at atmosphere <0.1 psig, flowrate 0.5 to 1.5 LPM. Vacuums to 12.5 psia are permissible. For In-situ glovebox and chambers (instrument or sensor in chamber) with environment between atmosphere and 1.5 psig. Vacuums to 12.5 psia are permissible. |
3000 with AC power 0-100/1000/10000 PPM no options |
Series 3000 Trace Oxygen Analyzer with three alarm relays, instrument status alarm relay, digital display, audible alarm, visual alarm indicators, and two manual sensor isolation valves. "Y" Auto Multi-range 0-100/1,000/10,000 PPM. Powered by universal AC (90-264VAC). Features a benchtop portable enclosure with carry handle. Includes standard connectors on the sample outlet, no filter. Also includes a standard 4-20mADC and 0-2vdc analog outputs. Warranty: 2-year electronics and 1-year sensor. No options included. No Options Application: For extractive gas sampling venting to atmosphere with sample gas pressure between 0.1 to 1.5 psig, flowrate 0.5 to 1.5 LPM or customer will supply their own gas sample conditioning options. |